OpenPolScope MantisBT

View Issue Details Jump to Notes ] [  ] Issue History ] Print ]
IDProjectCategoryView StatusDate SubmittedLast Update
0000087OpenPolScope-API[All Projects] bugpublic2019-01-07 12:122019-01-07 12:12
Reporterrudolfo 
Assigned Toamit 
PriorityhighSeveritymajorReproducibilityalways
StatusassignedResolutionopen 
PlatformNikon inverted, Ham Flash4OSWindowsOS Version
Product Version 
Target VersionFixed in Version 
Summary0000087: frame averaging with Flash4 crashes Pol-Acquisition plugin
DescriptionOn Friday, Jan 4, Mai and I used the Nikon inverted stand with the light field camera that now has the new Hamamatsu Flash4 camera with USB connection installed. The Pol-Acquisition plugin in polarized fluorescence mode worked without a hitch, except when we enabled frame averaging. With frame averaging enabled (4 frames), a time series acquisition stalled halfway, and MicroManager needed to be rebooted. Frame averaging seemed to work for some of the acquisitions, but then failed.

Following this experience, we disabled frame averaging (we only recorded sample images, no background was needed), and the rest of the session, taking about 6 hours with many (beautiful) data sets, went flawlessly.
TagsNo tags attached.
Attached Files

- Relationships

-  Notes
There are no notes attached to this issue.

- Issue History
Date Modified Username Field Change
2019-01-07 12:12 rudolfo New Issue
2019-01-07 12:12 rudolfo Status new => assigned
2019-01-07 12:12 rudolfo Assigned To => amit


Copyright © 2000 - 2026 MantisBT Team